| Other Options Include: |
| • |
Add connector. An additional connector of the same type may be ordered. This provides the user the ability to change out or replace connectors during development phases of a project. Once the device's connector has been removed and replaced with the same or different connector, the device's calibration and match are not guaranteed. |
| • Add male connector, option 20M |
| • Add female connector, option 20F |
| A connector can be changed in about 5 minutes. |
| • |
Add recorder out, option 001. This option adds an optional output voltage that is proportional to the input power. The recorder output voltage ranges from zero to one volt and is proportional to input power level. The range of power levels represented by the voltage output is set by the user. The output is provided through the SMB connector that is located in the same position as the trigger in. |
| If the unit is ordered with both external hardware triggering and the recorder out options, the SMB connector is shared between the recorder out and trigger in functions. Trigger in and recorder out functions are mutually exclusive functions. As a result, when the unit's recorder output is turned on the trigger input is disabled. |
| • |
Add best match, option 002. The PowerSensor+ products already have the best standard match in the industry. However, this still may be inadequate for some purposes. To meet these special needs, Ladybug provides a limited number of hand tuned units. The VSWR on these selected, hand tuned units is about1.05:1. Please call before ordering to obtain additional details. |
| • |
Add external hardware triggering, option 003. Adds an external hardware input trigger and external hardware output trigger. Both trigger in and trigger out are SMB connectors. If the recorder out option is also present, the trigger in SMB connector is shared between trigger in and recorder out (see "Add recorder out"). As a result, when the unit's trigger input is turned on the recorder output is disabled. |
| • |
Add additional hardware video filters, option 004. These filters are intended for pulse profiling. The LB480A comes standard with a 100 kHz hardware filter and additional lower frequency software filters. Option 004 offers four user selectable hardware filters: 300 kHz, 500 kHz, 1 MHz, and 10 MHz. The option also includes a number of automated pulse measurements such as rise time, fall time, droop, pulse width, overshoot and undershoot. |
| • |
Add real-time, calibrated detector output capability with option OW2. This capability allows viewing of one-shot pulses and other real-time modulated
signals by using a high speed oscilloscope to display the detector's analog output. The option OW2 signal is available on the TO SMB connector (trigger out signal
will not be available). Option OW2 requires option 004 and is not available when option 001 is ordered.
|